Character
● Single-chip microcomputer control, LCD Screen (128*64 Dots)
● Large LCD screen, can display multiple sets of data.
● Large memory space ,can store multiple sets of data and curve.
● Auto zero,100% adjustment function
● Auto wavelength adjustment
● Auto filter changing, large sample pool (5mm ~ 100mm)
● With up to ten standard sample for measurement function of standard curve establish
● Through direct input K, B factor is established for quantitative measurement of standard curve
● Can be directly input standard sample and the corresponding concentration values to establish the standard curve for quantitative measurement
● Switch off power save standard curve parameter measuring set
● Can be done any standard curve editor, convenient for use.
● With general parallel printer interface or printer interface, can print title bar, measurement data, curve of standard sample curve, and the curve.
● With USB interface
● Can achieve more accurate and flexible measurement with PC control (optional) Technical Specifications
Wavelength Range | 325~1000nm |
Band Width | 2nm |
Wavelength Accuracy | ±1.0nm |
Wavelength Repeatability | 0.2nm |
Photometric Accuracy | ±0.5% τ |
Photometric Repeatability | 0.2% τ |
Stray Light | ≤0.05% τ (340nm NaNO2) |
Stability | 0.001A/30min@500nm |
Photometric Mode | T,A,C,E |
Wavelength Setting | Automatic |
Photometric Display Range | -0.3~3A |
Display mode | LCD Screen (128*64 Dots) |
Detector | Import Silicon Photodiode Detector |
Light Source | Import Tungsten Lamp |
Power Requirement | AC 220V/50Hz 110V/60Hz |
Power | 120W |
Dimensions(W*D*H) | 460×330×210mm |
Weight | 11Kg |




|