Vertical / Horizontal Flame Apparatus Lab Testing Equipment 5V / HB/ V-0/1/2 UL94

Categories Electrical Safety Test Equipment
Brand Name: PEGO GROUP
Model Number: HVR-4H
Certification: Third-lab calibration certificate
Place of Origin: China
MOQ: 1 set
Price: USD3100-USD4500/set
Payment Terms: T/T, Western Union, MoneyGram
Delivery Time: 15 working days
Packaging Details: each packed in a wooden case
Reference Standard: UL94, IEC60695-11-4 and IEC60695-11-3
Production name: flammability test equipment
Application: V-0/1/2, HB, 5VA/B, HF-1/2 and HBF
Volume of chamber: 0.75CMB
Control units: single chip and button control
Dimension: 1100mm(W)*700mm(D)*1300mm (H)
Working power: single phase 220V/50-60Hz
Warranty: 12 months
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
  • Submit Buying Request
    • Product Details
    • Company Profile

    Vertical / Horizontal Flame Apparatus Lab Testing Equipment 5V / HB/ V-0/1/2 UL94

    5V / HB/ V-0/1/2 UL94 Vertical and Horizontal Flame Test Apparatus Lab Testing Equipment


    1. Introduction of Flame test Apparatus

    Horizontal and vertical flame test apparatus is applied to 50W vertical flame test includes V-0, V-1, V-2, horizontal flame tester HB, and 500W vertical flame test includes 5VA, 5VB, HF-1, HF-2 and HBF. It is widely used in the R&D department, QC department and production department of lighting equipment, low-voltage apparatus, household appliance, motor and etc.. And also can be used in the industries of insulation material, plastic, non-metalic material, foam with a density of not less than 250kg/m' and other solid combustible material.


    Horizontal and vertical flame test apparatus is applied to simulate the fire at the early stage surround the electric and electronics appliance, and then evaluate the degree of fire hazard. It is designed according to UL94, IEC60695-11-4 and IEC60695-11-3. The flammability test uses a bunsen burner of a specified size and specified gas source (methane or natural gas) to ignite the horizontal or vertical specimen for several times in a certain flame height and a certain flame angle, the fire hazard is evaluated by burning rate, after glow time, after flame time and damage length of the specimen.


    The horizontal and vertical flame test apparatus is controlled by single chip and push button. The test apparatus is built-in air exhaust, the exhaust system can be preset to delay 1-3mins after power off. The front panel is made by military-grade explosion-proof glass and transparent for easy observation of the experimental process, and the shell is made by coated hardness steel. Fully independent electrical circuit and gas circuit design for safer use.


    2. Parameter of Flame Test Apparatus

    BurnerΦ9.5mm ± 0.3mm, 100mm length bunsen burner, source: 50W and 500W
    Dip angle00, 200, 450 (manual adjustment)
    Flame height20mm ± 2mm to 180mm±10mm (adjustable), record by height gauge
    Flame time0-999.9s±0.1s (adjustable)
    After flame time0-999.9s±0.1s (auto record, manual pause)
    After glow time0-999.9s±0.1s (auto record, manual pause)
    Number of flame0~9999 times (presettable)
    Fuel gas98% methane or 37MJ/m3±1MJ/M3 natural gas
    Flow and pressure recordDouble flowmeters and manometer
    Gas Pressure0.1Mpa
    Gas Flow50W:105±5ml/min or 18±1 mm, 500W:965±30ml/mino or 117±3 mm
    Flame calibrations

    1). Flame temperature required: heat up from 1000C±50C to 7000C±30C within the time of 44s±2s (50W) and 54s±2s (500W)

    2). Flame height: 50W: 20±2mm (50W) 500W: 40±2mm (inner), 125±10mm (outer)

    3). Thermocouple: Φ0.5mm K type thermocouple (The insulating sleeve heat resist to 1250 ℃), 2 pieces to detect the temperature

    4). Temperature measuring copper blocks

    Test backgroundblack
    Volume of chamber0.75 CBM, exhaust hole: Φ70mm
    Thermal radiation baffleAutomatic masking
    Safe operation interfaceautomatically power off upon open the glass window
    AccessoriesPick-up tray, sample clamp, wrapping tissue 12g/m2 ~30g/m2, steel wire mesh, height gauge, Cotton and etc.
    Dimension of apparatus1100mm(W)*700mm(D)*1300mm (H)
    Power supplyAC220V, 50Hz, ≤500W

    3. FAQ

    1). What is the warranty period of the equipment?
    Answer: the equipment warranty is 12 months upon you receipt of the equipment


    2). Can you provide third-lab calibration certificate?
    Answer: Yes, the equipment can send to the third-labs authorized by ISO17025 and get the calibration certificate, but the buyer should undertake the charges.


    3). What is the working power of the equipment?
    Answer: single phase 220v/50-60Hz


    4). What is the language of the operation interface?

    Answer: English